High definition Emcee Conductivity Meter - Online Chlorine Dioxide Meter T4053 – Chunye Detail:
Function
Online chlorine dioxide meter is a microprocessor-based water quality online monitoring control instrument.
Typical Use
This instrument is widely used in online monitoring of water supply, tap water, rural drinking water, circulating water, washing film water, disinfectant water, pool water. and other industrial processes. It continuous monitoring and control chlorine dioxide and temperature value in aqueous solution.
Mains Supply
85~265VAC±10%,50±1Hz, power ≤3W;
9~36VDC, power consumption≤3W;
Measuring Range
Chlorine dioxide: 0~20mg/L; 0~20ppm;
Temperature: 0~150℃.
Features
1.Large display, standard 485 communication, with online and offline alarm, 98*98*130mm meter size, 92.5*92.5mm hole size, 3.0 inch large screen display.
2.Historical curve: The chlorine dioxide measurement data can be stored automatically every 5 minutes, and the residual chlorine value can be stored continuously for a month. Provide standard solution calibration and field calibration.
3.Built-in various measurement functions, one machine with multiple functions, meeting the requirements of various measurement standards.
4.The design of the whole machine is waterproof and dustproof, and the back cover of the connection terminal is added to extend the service life in harsh environments.
5.Panel/wall/pipe installation, three options are available to meet various industrial site installation requirements.
Electrical connections
Electrical connection The connection between the instrument and the sensor: the power supply, output signal, relay alarm contact and the connection between the sensor and the instrument are all inside the instrument. The length of the lead wire for the fixed electrode is usually 5-10 meters, and the corresponding label or color on the sensor Insert the wire into the corresponding terminal inside the instrument and tighten it.
Instrument installation method

Technical specifications
| Measurement range | 0.005~20.00mg/L ; 0.005~20.00ppm |
| Measurement unit | Potentiometric method |
| Resolution | 0.001mg/L ; 0.001ppm |
| Basic error | ±1%F.S
։ ˫ |
| Temperature | -10 150.0 ( Based on sensor)
˫ |
| Temperature Resolution | 0.1
˫ |
| Temperature Basic error | ±0.3
։ |
| Current output | 2 groups: 4 20mA |
| Signal output | RS485 Modbus RTU |
| Other functions | Data record &Curve display |
| Three relay control contacts | 2 groups:5A 250VAC,5A 30VDC |
| Optional power supply | 85~265VAC,9~36VDC,power consumption≤3W |
| Working conditions | No strong magnetic field interference around except the geomagnetic field.
։ ˫ |
| Working temperature | -10 60 |
| Relative humidity | ≤90% |
| Waterproof rating | IP65 |
| Weight | 0.6kg |
| Dimensions | 98×98×130mm |
| Installation opening size | 92.5×92.5mm |
| Installation methods | Panel & wall mounted or pipeline |

|
Model No. |
CS5560 |
|
Measurement method |
Tri-electrode method |
|
Measure material |
Double liquid junction,annular liquid junction |
|
Housing material/Dimensions |
PP, Glass, 120mm*Φ12.7mm |
|
Waterproof grade |
IP68 |
|
Measurement range |
0 – 5.000 mg/L, 0 – 20.00 mg/L |
|
Accuracy |
±0.05mg/L; |
|
Pressure resistance |
≤0.3Mpa |
|
Temperature compensation |
None or Customize NTC10K |
|
Temperature range |
0-50℃ |
|
Calibration |
Sample calibration |
|
Connection methods |
4 core cable |
|
Cable length |
Standard 5m cable, can be extended to 100m |
|
Installation thread |
PG13.5 |
|
Application |
Tap water, disinfectant fluid, etc. |
Product detail pictures:
Related Product Guide:
Our organization has been focusing on brand strategy. Customers' gratification is our greatest advertising. We also source OEM provider for High definition Emcee Conductivity Meter - Online Chlorine Dioxide Meter T4053 – Chunye , The product will supply to all over the world, such as: kazakhstan, Estonia, Oman, Our company has built stable business relationships with many well-known domestic companies as well as oversea customers. With the goal of providing high quality products to customers at low cots, we are committed to improving its capacities in research, development, manufacturing and management. We have honored to receive recognition from our customers. Till now we have passed ISO9001 in 2005 and ISO/TS16949 in 2008. Enterprises of "quality of survival, the credibility of development" for the purpose, sincerely welcome domestic and foreign businessmen to visit to discuss cooperation.