Massive Selection for Water Quality On Line Monitor - T9000 CODcr Water Quality On-line Automatic Monitor – Chunye Detail:
T9000 CODcr Water Quality On-line Automatic Monitor
Product Principle
Water samples, potassium dichromate digestion solution, silver sulfate solution (silver sulfate as catalyst can be added to oxidize linear aliphatic compounds more effectively) and concentrated sulfuric acid mixture heated to 175℃. The color of organic compounds in dichromate ion oxidation solution will change. The analyzer detects the color change and converts the change into COD value then output the value.The amount of dichromate ion consumed is equivalent to the amount of oxidizable organic matter, namely COD.
Technical Parameters:
|
No. |
Name |
Technical Specifications |
|
1 |
Application Range |
Suitable for wastewater with COD in the range of 10~5,000mg/L and chloride concentration less than 2.5g/L Cl-. According to customers’ actual demand, it can be extended to wastewater with chloride concentration less than 20g/L Cl-. |
|
2 |
Test Methods |
Potassium dichromate digestion at high temperature, colorimetric determination |
|
3 |
Measuring range |
10~5,000mg/L |
|
4 |
Lower limit of Detection |
3 |
|
5 |
Resolution |
0.1 |
|
6 |
Accuracy |
±10% or ±8mg/L(Take the bigger value) |
|
7 |
Repeatability |
10% or 6mg/L(Take the larger value) |
|
8 |
Zero Drift |
±5mg/L |
|
9 |
Span Drift |
±10% |
|
10 |
Measurement cycle |
Minimum 20 minutes. According to the actual water sample, the digestion time can be set from 5 to120 minutes. |
|
11 |
Sampling period |
Time interval (adjustable), integral hour or trigger measurement mode can be set. |
|
12 |
Calibration cycle |
Automatic calibration (1-99 days adjustable), according to actual water samples, manual calibration can be set. |
|
13 |
Maintenance cycle |
Maintenance interval is more than one month, about 30 minutes each time. |
|
14 |
Human-machine operation |
Touch screen display and instruction input. |
|
15 |
Self checking protection |
Working status is self-diagnostic, abnormal or power failure will not lose data. Automatically eliminates residual reactants and resumes work after abnormal reset or power failure. |
|
16 |
Data storage |
No less than half a year data storage |
|
17 |
Input interface |
Switch quantity |
|
18 |
Output interface |
Two RS485 digital output, One 4-20mA analog output |
|
19 |
Working Conditions |
Working indoors; temperature 5-28℃; relative humidity≤90% (no condensation,no dew) |
|
20 |
Power Supply Consumption |
AC230±10%V, 50~60Hz, 5A |
|
21 |
Dimensions |
355×400×600(mm) |
Product detail pictures:
Related Product Guide:
Adhering into the theory of "quality, services, efficiency and growth", now we have gained trusts and praises from domestic and international shopper for Massive Selection for Water Quality On Line Monitor - T9000 CODcr Water Quality On-line Automatic Monitor – Chunye , The product will supply to all over the world, such as: Bahamas, Germany, Bangladesh, Our next goal is to exceed the expectations of every client by offering outstanding customer service, increased flexibility and greater value. All in all, without our customers we do not exist; without happy and fully satisfied customers, we fail. We are looking for the wholesale, Drop ship. Please contact us if you are interesting our products. Hope to do business with you all. High quality and fast shipment!
In China, we have many partners, this company is the most satisfying to us, reliable quality and good credit, it is worth appreciation.