Product Overview:
The urea online monitor uses spectrophotometry for detection. This instrument is mainly used for online monitoring of swimming pool water.
This analyzer can operate automatically and continuously without human intervention for a long time based on on-site settings, and is widely applicable for online automatic monitoring of urea indicators in swimming pools.
Product principle:
Urea reacts with diacetylone and antipyrine to produce a yellow color, and its absorbance is proportional to the content of urea.
technical specification
|
Number |
Specification Name |
Technical specification parameters |
|
1 |
test method |
Diacetyl oxime spectrophotometric method |
|
2 |
measuring span |
0~10mg/L(Segmented measurement, with automatic switching capability) |
|
3 |
lower limit of detection |
0.05 |
|
4 |
Resolution |
0.001 |
|
5 |
Accuracy |
±10% |
|
6 |
Repeatability |
≤5% |
|
7 |
zero drift |
±5% |
|
8 |
span drift |
±5% |
|
9 |
measurement period |
Less than 40 minutes, the dissolution time can be set. |
|
10 |
sampling period |
Time interval (adjustable), hourly or trigger measurement mode, can be set |
|
11 |
calibrating period |
Automatic calibration (adjustable from 1 to 99 days), and manual calibration can be set based on actual water samples. |
|
12 |
maintenance period |
The maintenance interval is more than 1 month, and each time it lasts approximately 5 minutes. |
|
13 |
Human-machine operation |
Touchscreen display and command input |
|
14 |
Self-check protection |
The instrument has a self-diagnosis function for its working status. Even if there is an anomaly or a power failure, the data will not be lost. In case of an abnormal reset or a power failure followed by a power restoration, the instrument will automatically remove the remaining reactants and automatically resume operation. |
|
15 |
data storage |
5-year data storage |
|
16 |
One-click maintenance |
Automatically empty the old reagents and clean the pipelines; replace the new reagents, automatically calibrate and automatically verify; it can also be selected to automatically clean the digestion chamber and metering tube using cleaning solution. |
|
17 |
Quick debugging |
Realize unmanned operation, continuous operation, and automatic generation of debugging reports, which greatly facilitates users and reduces labor costs. |
|
18 |
input interface |
switching value |
|
19 |
output interface |
1 RS232 output, 1 RS485 output, 1 4-20mA output |
|
20 |
work environment |
For indoor work, the recommended temperature range is 5 to 28 degrees Celsius, and the humidity should be no more than 90% (without condensation). |
|
21 |
Power supply |
AC220±10%V |
|
22 |
Frequency |
50±0.5Hz |
|
23 |
Power |
≤150W,Without sampling pump |
|
24 |
Inches |
Height: 520 mm, Width: 370 mm, Depth: 265 mm |