Online Automatic Monitoring Instrument for Orthophosphate Water Quality

Short Description:

Phosphorus Hazards to Marine Life Mostmarine organisms are highly sensitive to organophosphorus pesticides. Concentrations that cause no reaction in pesticide-resistant insects can rapidly prove lethal to marine life. The human body contains an essential neurotransmitter enzyme called acetylcholinesterase.


Product Detail

Product Tags

Product Description

 

Phosphorus Hazards to Marine Life Most

marine organisms are highly sensitive to

organophosphorus pesticides. Concentrations

that cause no reaction in pesticide-resistant

insects can rapidly prove lethal to marine

life. The human body contains an essential

neurotransmitter enzyme called

acetylcholinesterase.

Organophosphorus compounds inhibit this enzyme, preventing it from breaking down acetylcholine. This leads to the accumulation of acetylcholine in the nervous system, causing poisoning and potentially fatal outcomes in severe cases. Long-term exposure to low doses of organophosphate pesticides can cause chronic poisoning and may pose carcinogenic and teratogenic risks to humans.

 

Product Principle

The water sample, catalyst solution, and strong oxidizing agent digestion solution are mixed. Under high-temperature, high-pressure acidic conditions, polyphosphates and other phosphorus-containing compounds in the water sample are oxidized by the strong oxidizing agent to form phosphate ions. In the presence of the catalyst, these phosphate ions react with the molybdate-containing strong acid solution to form a colored complex. The analyzer detects this color change and converts it into an orthophosphate value output. The amount of colored complex formed corresponds to the orthophosphate content.

 

Technical Specifications

SN

Specification Name

Technical Specifications

1

Test Method

Phosphomolybdenum Blue Spectrophotometric Method

2

Measurement Range

0–50 mg/L (segmented measurement, expandable)

3

Accuracy

20% of full scale standard solution, not exceeding ±5%

50% of full scale standard solution, not exceeding ±5%

80% of full scale standard solution, not exceeding ±5%

4

Limit of Quantification

0.02mg/L

5

Repeatability

2%

6

24h Low-Concentration Drift

0.01mg/L

7

24h High-Concentration Bleaching

1%

8

Measurement Cycle

Minimum test cycle: 20 minutes, configurable

9

Sampling Cycle

Time interval (adjustable), hourly, or trigger measurement mode, configurable

10

Calibration Cycle

Auto-calibration (adjustable from 1 to 99 days), manual calibration can be set based on actual water samples.

11

Maintenance Cycle

Maintenance intervals exceed one month, with each session lasting approximately 5 minutes.

12

Human-Machine Operation

Touchscreen Display and Command Input

13

Self-Diagnostic Protection

The instrument performs self-diagnostics during operation and retains data after abnormalities or power loss. Following abnormal resets or power restoration, it automatically purges residual reagents and resumes normal operation.

14

Data Storage

5-Year Data Storage

15

One-Button Maintenance

Automatically drains old reagents and cleans tubing; replaces new reagents, performs automatic calibration and verification; optional automatic cleaning of digestion cells and metering tubes with cleaning solution.

16

Quick Debugging

Achieve unattended, uninterrupted operation with automatic generation of debugging reports, greatly enhancing user convenience and reducing labor costs.

17

Input Interface

switching value

18

Output Interface

1 channel RS232 output, 1 channel RS485 output, 1 channel 4–20 mA output

19

Operating Environment

Indoor operation, recommended temperature range: 528, humidity 90% (non-condensing)

20

Power Supply

AC220±10%V

21

Frequency

50±0.5Hz

22

Power

≤150 W (excluding sampling pump)

23

Dimensions

520 mm (H) × 370 mm (W) × 265 mm (D)

 


  • Previous: