Product Description:
Industries such as electroplating, chemical
processing, textile dyeing, battery manufacturing,
and metal fabrication generate zinc-containing
wastewater. Excessive zinc is harmful to human
health and may even pose carcinogenic
risks. Furthermore, using zinc-contaminated
wastewater for agricultural irrigation severely
impairs crop growth, particularly wheat. Excess zinc deactivates enzymes in soil, weakens microbial biological functions, and ultimately impacts human health through the food chain.
Product Principle:
This product employs spectrophotometric colorimetry for determination. After mixing the water sample with the conditioning agent, zinc in all forms is converted into zinc ions. In an alkaline environment and in the presence of a sensitizer, these zinc ions react with an indicator to form a colored complex. The analyzer detects this color change and converts it into a zinc value for output. The amount of colored complex formed corresponds to the zinc content.
Technical Specifications:
|
SN |
Specification Name |
Technical Specifications |
|
1 |
Test Method |
Zinc Reagent Colorimetric Method |
|
2 |
Measurement Range |
0–30 mg/L (segmented measurement, expandable) |
|
3 |
Detection limit |
≤0.02 |
|
4 |
Resolution |
0.001 |
|
5 |
Accuracy |
±10% |
|
6 |
Repeatability |
≤5% |
|
7 |
Zero drift |
±5% |
|
8 |
Range drift |
±5% |
|
9 |
Measurement Cycle |
Minimum test cycle: 30 minutes, configurable |
|
10 |
Sampling Cycle |
Time interval (adjustable), hourly, or trigger measurement mode, configurable |
|
11 |
Calibration Cycle |
Auto-calibration (adjustable from 1 to 99 days), manual calibration can be set based on actual water samples. |
|
12 |
Maintenance Cycle |
Maintenance intervals exceed one month, with each session lasting approximately 5 minutes. |
|
13 |
Human-Machine Operation |
Touchscreen Display and Command Input |
|
14 |
Self-Diagnostic Protection |
The instrument performs self-diagnostics during operation and retains data after abnormalities or power loss. Following abnormal resets or power restoration, it automatically purges residual reagents and resumes normal operation. |
|
15 |
Data Storage |
5-Year Data Storage |
|
16 |
One-Button Maintenance |
Automatically drains old reagents and cleans tubing; replaces new reagents, performs automatic calibration and verification; optional automatic cleaning of digestion cells and metering tubes with cleaning solution. |
|
17 |
Quick Debugging |
Achieve unattended, uninterrupted operation with automatic generation of debugging reports, greatly enhancing user convenience and reducing labor costs. |
|
18 |
Input Interface |
switching value |
|
19 |
Output Interface |
1 channel RS232 output, 1 channel RS485 output, 1 channel 4–20 mA output |
|
20 |
Operating Environment |
Indoor operation, recommended temperature range: 5–28℃, humidity ≤90% (non-condensing) |
|
21 |
Power Supply |
AC220±10%V |
|
22 |
Frequency |
50±0.5Hz |
|
23 |
Power |
≤150 W (excluding sampling pump) |
|
24 |
Dimensions |
1,470 mm (H) × 500 mm (W) × 400 mm (D) |