The Biological Toxicity Water Quality Online Monitor represents a transformative approach to water safety assessment by continuously measuring the integrated toxic effect of pollutants on living organisms, rather than merely quantifying specific chemical concentrations. This holistic biomonitoring system is critical for early warning of accidental or intentional contamination in drinking water sources, wastewater treatment plant influents/effluents, industrial discharges, and receiving water bodies. It detects synergistic effects of complex contaminant mixtures—including heavy metals, pesticides, industrial chemicals, and emerging pollutants—that conventional chemical analyzers may miss.By providing a direct, functional measure of water's biological impact, this monitor serves as an indispensable sentinel for protecting public health and aquatic ecosystems. It enables water utilities and industries to trigger immediate responses—such as diverting contaminated inflows, adjusting treatment processes, or issuing public alerts—long before traditional lab results are available. The system is increasingly integrated into smart water management networks, forming a vital component of comprehensive source water protection and regulatory compliance strategies in an era of complex pollution challenges.


Product Detail

Product Tags

Technical Specifications:

1.Measurement principle: Luminescent bacteria method

2.Bacterial working temperature: 15-20 degrees

3.Bacterial culture time: < 5 minutes

4.Measurement cycle: Fast mode: 5 minutes; Normal mode: 15 minutes; Slow mode: 30 minutes

5.Measurement range: Relative luminescence (inhibition rate) 0-100%, toxicity level

6.Temperature control error

(1) The system has an in-built integrated temperature control system (not external), with an error of ≤ ±2℃;

(2) The temperature control error of the measurement and culture chamber ≤ ±2℃;

(3) The temperature control error of the bacterial strain low-temperature preservation component ≤ ±2℃;

7.Reproducibility: ≤ 10%

8.Accuracy: Pure water detection light loss ± 10%, actual water sample ≤ 20%

9.Quality control function: Includes negative quality control, positive quality control and reaction time quality control; Positive quality control: 2.0 mg/L Zn2+ reaction for 15 minutes, inhibition rate 20%-80%; Negative quality control: Pure water reaction for 15 minutes, 0.6 ≤ Cf ≤ 1.8;

10.Communication port: RS-232/485, RJ45 and (4-20) mA output

11.Control signal: 2-channel switch output and 2-channel switch input; Supports linkage with sampler for over-limit retention function, pump linkage;

12.Has the function of automatic bacterial solution preparation, automatic bacterial solution usage days alarm function, reducing maintenance workload;

13. Has the function of automatic temperature alarm for detection and culture temperature;

14. Environmental requirements: Moisture-proof, dust-proof, temperature: 5-33℃;

15. Instrument size: 600mm * 600mm * 1600mm

16. Uses 10-inch TFT, Cortex-A53, 4-core CPU as the core, high-performance embedded integrated touch screen;

17. Other aspects: Has the function of recording the instrument operation process log; Can store at least one year of original data and operation logs; Instrument abnormal alarm (including fault alarm, over-range alarm, over-limit alarm, reagent shortage alarm, etc.); Data is automatically saved in case of power failure; TFT true-color liquid crystal touch screen display and command input; Abnormal reset and automatic recovery of working state after power failure and power restoration; Instrument status (such as measurement, idle, fault, maintenance, etc.) display function; The instrument has three-level management authority.