CS6603CD Digital RS485 Output COD BOD TOC TURBIDITY Sensor

Short Description:

Digital RS485 Output COD Sensor is an advanced online instrument for continuous, real‑time measurement of Chemical Oxygen Demand (COD) in water and wastewater. Unlike conventional COD analyzers that rely on wet chemistry (dichromate digestion, colorimetry), this optical sensor uses UV‑vis absorption spectroscopy at multiple wavelengths (e.g., 254 nm for organic matter, 550 nm for turbidity compensation) to estimate COD without reagents or sample digestion. The sensor integrates a xenon or LED light source, a photodetector array, and a built‑in temperature sensor for automatic compensation (ATC). It directly outputs COD values in mg/L via RS‑485 (Modbus RTU), enabling plug‑and‑play integration into SCADA or IoT platforms. Its rugged, waterproof (IP68) housing and low‑maintenance design (no pumps, no tubing, no reagents) are suitable for continuous submersed operation in wastewater influent/effluent channels, industrial discharge, and surface water monitoring. Typical COD range is 0–2000 mg/L, with response time of < 30 seconds. While optical COD sensors require site‑specific calibration against standard dichromate methods, they provide rapid trend data and early warning of organic pollution spikes, helping operators optimize aeration, reduce chemical usage, and prevent permit violations. The digital RS485 output ensures long‑distance data transmission (up to 1200 m) without signal degradation, supports multi‑sensor networking (daisy‑chain), and enables remote calibration and self‑diagnostics. Compared to conventional COD analyzers, this digital sensor eliminates the need for expensive reagents, hazardous waste disposal, and frequent maintenance, significantly reducing operational costs and environmental footprint, making it an essential tool for unattended, continuous water quality monitoring in both industrial and municipal applications. It is particularly suitable for applications where rapid trend detection and low maintenance are prioritized over laboratory‑grade absolute accuracy, such as early warning systems, process control in biological treatment, and pollutant load estimation.


Product Detail

Product Tags

Testing principle:

Many organic compounds dissolved in water are absorbent  to  ultraviolet  light. Therefore, the total amount of organic pollutants in the water can be measured by measuring the extent to which these organics absorb ultraviolet light at  254nm.

Sensor features: 

Digital sensor, RS-485 output, support Modbus

No reagent, no pollution, more economic and environmental protection Automatic compensation of turbidity interference, with excellent test performance

With self-cleaning brush, can prevent biological attachment, maintenance cycle more 

Technical parameters

Name Parameter
Interface Support RS-485, MODBUS protocols
COD Range 0.1 to 1500mg/L equiv.KHP
BOD Range 0.1 to 900mg/L equiv.KHP
COD/BOD Accuracy <5% equiv.KHP
COD/BOD Resolution 0.01mg/L equiv.KHP
TOC Range 0.1 to 750mg/L equiv.KHP
TOC Accuracy <5% equiv.KHP
TOC Resolution 0.1mg/L equiv.KHP
Tur Range 0.1-4000 NTU
Tur Accuracy <3% or 0.2NTU
Tur Resolution 0.1NTU
Temperature Range +5 ~ 45℃
Housing IP Rating IP68
Maximum pressure 1 bar
User Calibration one or two points
Power Requirements DC 12V +/-5%,current<50mA(without wiper)
Sensor OD 32 mm
Sensor Length 200 mm
Cable Length 10m (default)