Online Oil-in-Water Analyzer is a specialized instrument designed for the continuous, real-time monitoring of oil and hydrocarbon concentrations in water. It plays a critical role in environmental protection, industrial wastewater treatment, produced water monitoring in oil and gas operations, marine vessel ballast water management, and cooling water systems in power plants and refineries. By providing immediate detection of oil contamination, it enables rapid response to prevent environmental violations, protect aquatic ecosystems, and ensure regulatory compliance with stringent discharge limits (such as the 15 ppm standard for produced water).
The analyzer employs advanced optical detection technologies, most commonly ultraviolet fluorescence (UVF) or infrared absorption (IR) methods. UVF technology is particularly effective as it excites oil molecules with specific wavelengths of light, causing them to emit fluorescence proportional to the oil concentration. This method offers high sensitivity, fast response, and minimal interference from water itself. Some systems utilize laser-induced fluorescence or scattered light measurements for enhanced accuracy in challenging matrices like high turbidity or varying oil types.


Product Detail

Product Tags

Online Oil-in-water Analyzer SC8000OIL 

Instrument features:

    • LCD large screen liquid crystal display
    • Smart menu
    • Historical data records
    • Manual or automatic temperature compensation
    • three relay control switch
    • Upper limit, lower limit, and hysteresis control
    • 4-20mA & RS485 with multiple output options
    • The same interface displays input values, temperature, current values, etc.
    • Set a password to prevent unauthorized access

    Technical parameter

    (1) Measurement range (based on the sensor's range):

    Concentration: 0.1200 mg/L

    temperature -10150.0℃;

    (2) Resolution:

    Concentration: 0.1/1 mg/L;

    temperature 0.1℃;

    (3) Basic error:

    Concentration: ±5-10% (depending on the sensor range);

    temperature :±0.3℃;

    (4) Two-channel current output:

    0/4 to 20mA (with a load resistance of <750Ω);

    20-4mA (load resistance <750Ω);

    (5) Communication output: RS485 MODBUS RTU;

    (6) Three sets of relay control contacts:

    5A 250VAC5A 30VDC

    (7) Power Supply (Optional):

    85265VAC±10%,50±1Hz, power3W;

    9-36VDC, power: 3W;

    (8) Dimensions: 144×144×118 mm;

    (9) Installation methods: panel-mounted, wall-mounted, or duct-mounted;

    Panel opening size: 137×137mm;

    (10) Protection rating: IP65;

    (11) Instrument weight: 0.8 kg;

    (12) Instrument operating environment:

    Environmental temperature: -10 to 60°C;

    Relative humidity: not greater than 90%;

    There is no strong magnetic field interference except the earth magnetic field.