T4077 Suspended solids (sludge) concentration meter

Short Description:

Water treatment plant (sedimentation tank), paper mill (pulp concentration), coal washing plant (sedimentation tank), power plant (ash slurry sedimentation tank), wastewater treatment plant (inlet/outlet, aeration tank, return sludge, primary sedimentation tank, secondary sedimentation tank, thickening tank, sludge dewatering)suspended solids (sludge) concentration meter is an online instrument used to continuously measure the concentration of suspended particles in liquids, such as activated sludge in wastewater treatment. It typically employs an optical backscatter or infrared absorption principle, where a light beam is directed into the sample, and the scattered or transmitted light is measured to determine solids concentration. The meter features automatic cleaning (e.g., wiper or air blast) to prevent fouling, automatic temperature compensation, and a wide measurement range (e.g., 0–50,000 mg/L or up to 200,000 mg/L for sludge). With analog (4-20mA) and digital (RS-485 Modbus) outputs, it integrates easily into control systems, enabling real‑time process optimization in aeration tanks, settling basins, and dewatering units, improving efficiency and regulatory compliance.


Product Detail

Product Tags

Typical Application:

Water treatment plant (sedimentation tank), paper mill (pulp concentration), coal washing plant (sedimentation tank), power plant (ash slurry sedimentation tank), wastewater treatment plant (inlet/outlet, aeration tank, return sludge, primary sedimentation tank, secondary sedimentation tank, thickening tank, sludge dewatering)

Instrument Features:

  • LCD color liquid crystal display
  • Smart Menu Operations
  • Manual or automatic temperature compensation
  • Three relay-controlled switches
  • Upper limit, lower limit, hysteresis control
  • 4-20 mA & RS485: Multiple output options
  • The same interface displays concentration values, temperature, current values, etc.
  • A password can be set to prevent unauthorized personnel from making accidental operations.

Technical parameter :

(1)Measurement range (based on the sensor range):

(2)0~200000mg/L;

(2) Measurement units: mg/L, g/L;

(3) Resolution: 0.001/0.01/0.1/1;

(4) Dual-channel current output:

0.4–20 mA (load resistance <750 Ω);

20–4 mA (load resistance <750 Ω);

(5) Communication output: RS485 MODBUS RTU;

(6) Two sets of relay control contacts:

3A 250VAC,3A 30VDC;

(7) Power supply (optional):

85–265 VAC ±10%,50 ±1 Hz, power ≤3 W;

9–36 VDC, power: ≤3 W;

(8) Dimensions: 98×98×130 mm;

(9) Installation methods: panel-type, wall-mounted;

Panel opening dimensions: 92.5×92.5 mm;

(10) Protection rating: IP65;

(11) Instrument weight: 0.6 kg;

(12) Instrument operating environment:

Environmental temperature: -10 to 60°C;

Relative humidity: no more than 90%;

There are no strong magnetic field disturbances around it other than the Earth's magnetic field.